We measure photon-number statistics of pulsed fields containing mesoscopic photon numbers in a time duration shorter than any feasible measuring time. As the detector we use a photomultiplier tube delivering anodic outputs in whose pulse-height spectrum only the zero detected-photon peak is resolved. We demonstrate that we can reconstruct the correct detected-photon statistical distribution provided the maximum number of detected photons falls within the range of linear response of the photomultiplier tube.
Photon-number statistics measured with a noncounting PMT
Andreoni A.
;Allevi A.;
2009-01-01
Abstract
We measure photon-number statistics of pulsed fields containing mesoscopic photon numbers in a time duration shorter than any feasible measuring time. As the detector we use a photomultiplier tube delivering anodic outputs in whose pulse-height spectrum only the zero detected-photon peak is resolved. We demonstrate that we can reconstruct the correct detected-photon statistical distribution provided the maximum number of detected photons falls within the range of linear response of the photomultiplier tube.File | Dimensione | Formato | |
---|---|---|---|