Infants <28 weeks' gestation in need of inflations at birth were recorded with Respiratory Function Monitor. Two devices were used for resuscitation. Peak Inspiratory Pressure spikes were visible in all inflations with GE Panda and in none with Neo-Puff. There was no significant difference in mean Vte/kg between GE Panda and Neo-Puff.
Pressure and tidal volume delivery in extremely preterm infants at birth using different t-piece resuscitation devices
Bresesti I.;
2023-01-01
Abstract
Infants <28 weeks' gestation in need of inflations at birth were recorded with Respiratory Function Monitor. Two devices were used for resuscitation. Peak Inspiratory Pressure spikes were visible in all inflations with GE Panda and in none with Neo-Puff. There was no significant difference in mean Vte/kg between GE Panda and Neo-Puff.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.