Infants <28 weeks' gestation in need of inflations at birth were recorded with Respiratory Function Monitor. Two devices were used for resuscitation. Peak Inspiratory Pressure spikes were visible in all inflations with GE Panda and in none with Neo-Puff. There was no significant difference in mean Vte/kg between GE Panda and Neo-Puff.

Pressure and tidal volume delivery in extremely preterm infants at birth using different t-piece resuscitation devices

Bresesti I.;
2023-01-01

Abstract

Infants <28 weeks' gestation in need of inflations at birth were recorded with Respiratory Function Monitor. Two devices were used for resuscitation. Peak Inspiratory Pressure spikes were visible in all inflations with GE Panda and in none with Neo-Puff. There was no significant difference in mean Vte/kg between GE Panda and Neo-Puff.
2023
Preterm infants; Respiratory function monitor; Resuscitation; T-piece resuscitator
Bresesti, I.; Cavigioli, F.; Scrivens, A.; Lista, G.; Chaban, B.; Roehr, C. C.; Zivanovic, S.
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11383/2176114
 Attenzione

L'Ateneo sottopone a validazione solo i file PDF allegati

Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 3
  • ???jsp.display-item.citation.isi??? 3
social impact