This study presents a scanning electron microscopy analysis of a failed PEEK retainer in an orthodontic patient. After 15 months of use, the patient reported a gap opening between teeth 41 and 42. The PEEK retainer was removed and sent for electron microscope analysis. To investigate the failure, scanning electron microscopy was employed to assess the microstructure and composition of the retainer at various magnifications. These findings suggest that the failure of the PEEK retainer was multifaceted, implicating factors such as material defects, manufacturing flaws, inadequate design, environmental factors, and patient-related factors. In conclusion, this scanning electron microscopy analysis offers valuable insights into the failure mechanisms of PEEK retainers in orthodontic applications. Further research is necessary to explore preventive strategies and optimize the design and fabrication of PEEK retainers, minimizing the occurrence of failures in orthodontic practice.

Failed Orthodontic PEEK Retainer: A Scanning Electron Microscopy Analysis and a Possible Failure Mechanism in a Case Report

Zecca P. A.
Primo
Conceptualization
;
Carganico A.
;
Borgese M.;Levrini L.;Reguzzoni M.
Ultimo
Data Curation
2024-01-01

Abstract

This study presents a scanning electron microscopy analysis of a failed PEEK retainer in an orthodontic patient. After 15 months of use, the patient reported a gap opening between teeth 41 and 42. The PEEK retainer was removed and sent for electron microscope analysis. To investigate the failure, scanning electron microscopy was employed to assess the microstructure and composition of the retainer at various magnifications. These findings suggest that the failure of the PEEK retainer was multifaceted, implicating factors such as material defects, manufacturing flaws, inadequate design, environmental factors, and patient-related factors. In conclusion, this scanning electron microscopy analysis offers valuable insights into the failure mechanisms of PEEK retainers in orthodontic applications. Further research is necessary to explore preventive strategies and optimize the design and fabrication of PEEK retainers, minimizing the occurrence of failures in orthodontic practice.
2024
2024
orthodontic relapse; orthodontic retainer; polyetheretherketone; scanning electron microscopy SEM
Zecca, P. A.; Bocchieri, S.; Carganico, A.; Caccia, M.; Fastuca, R.; Borgese, M.; Levrini, L.; Reguzzoni, M.
File in questo prodotto:
File Dimensione Formato  
dentistry PeeK-12-00223.pdf

accesso aperto

Tipologia: Versione Editoriale (PDF)
Licenza: Creative commons
Dimensione 2.16 MB
Formato Adobe PDF
2.16 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/11383/2177551
 Attenzione

L'Ateneo sottopone a validazione solo i file PDF allegati

Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 1
social impact